Skip to content

Evaluate experimental data

Use the measured image and the fitted spectra and abundances to see where unmixing leaves signal unexplained. These diagnostics work without knowing which components are present or how much of each is in a pixel.

To reproduce the plots, run Estimate unknown spectra through Refine the spectra with MCR-ALS. The walkthrough uses the simulated mixture as a stand-in for an experimental image. Its ground truth is excluded from the calculations below. For your own data, use the unfolded measured image and its paired fitted abundances and spectra.

Map reconstruction errors

Multiply the fitted abundances by their spectra to predict the detector signal at every pixel. Fold the prediction back into the image. This example uses YXC arrays, with the detector channel axis last as required by reconstruction_error_maps.

from lumic.evaluation import reconstruction_error_maps

observed = unfolded.fold_spectra().data
reconstructed = unfolded.fold_spectra(refined_counts @ refined_spectra).data
errors = reconstruction_error_maps(observed, reconstructed)
  • Signed residuals (errors.residuals) are measured minus predicted signal, one value per pixel and channel. Positive values show where the model predicted too little signal.
  • RMS error (errors.rms) gives the size of the residual across channels, in detector intensity units. Use it to locate large mismatches. Shot noise can produce larger absolute errors in bright regions.
  • Relative L2 error (errors.relative_l2) divides each pixel's residual norm by its measured spectral norm. Use it to compare mismatch relative to signal strength. It can be unstable near zero signal. Set min_signal to a threshold appropriate for your data and use errors.relative_valid to mask excluded pixels.

Look for structure in the residuals

A single RMS value per pixel hides which channels account for the mismatch. Residual PCA finds patterns across the signed channel residuals. Plot each PC's loadings against detector wavelength and its scores as an image. Spatially coherent scores can point to a region worth inspecting.

from lumic.evaluation import (
    residual_neighbor_correlation,
    residual_pca_from_residuals,
    residual_spectral_correlation,
)

pca = residual_pca_from_residuals(errors.residuals, n_components=3)
pc1_map = pca.scores[..., 0]
pc1_spectrum = pca.loadings[0]
mean_residual = pca.mean_residual
print(f"PC1 explains {pca.explained_variance_ratio[0]:.1%} of centered residual variance")

neighbor_correlation = residual_neighbor_correlation(errors.residuals)
channel_correlation_map = residual_spectral_correlation(errors.residuals)
print(f"Residual neighbor correlation: {neighbor_correlation:.3f}")

Inspect the mean residual for persistent channel bias that centered PCA removes. Neighbor correlation summarizes how similarly adjacent pixels behave across the image. Spectral correlation maps how similarly adjacent channels behave at each pixel. Inspect these alongside the error magnitude. Noise, fitting, detector artifacts, or a missing spectral component can all create structure. Use the diagnostics to guide inspection. They do not identify a fluorophore or establish statistical significance.

RMS error, relative L2 error, spatial PC1 scores, and the corresponding spectral loadings for the ternary image.

The maps show the signal region from the example. The PC1 score map and loading curve describe the same pattern in space and across channels.

Compare two fits without ground truth

Pass two sets of fitted spectra to compare_spectra to see how much each spectral shape changes. By default it pairs the components before comparing them. You can use this to check sensitivity to initialization or pixel selection. Stability across runs describes agreement between estimates.

from lumic.evaluation import compare_spectra

change = compare_spectra(initial_spectra, refined_spectra)
print("Spectral shape distances:", change.shape_distance)
print("Largest channel changes (percentage points):", 100 * change.max_channel_change)

shape_distance compares unit-L2 spectral shapes. max_channel_change compares the fraction of total signal assigned to a single channel after scaling each spectrum to sum to one.

If you have known spectra and abundance maps, you can evaluate recovery against simulation ground truth.

What does a missing component look like?

Add a narrow line of mNeptune to the noisy ternary example. The line contributes 150 expected counts per pixel on top of the original 500. Fit the resulting image with three components using the same blind workflow. The fit receives the measured pixels and the component count. The true location of the fourth fluorophore is shown in the figure for comparison.

from lumic import SpectralImage, essential_spectral_pixles, mcr_als
from lumic import positive_least_squares, unique_profiles
from lumic.simulation import (
    add_confocal_noise, load_offline_emission_spectrum, simulate_linear_mixture,
)

mneptune = detector.sample(*load_offline_emission_spectrum("mNeptune"))
mneptune = mneptune / mneptune.sum()
line = inside & (np.abs(y - 57) <= 1) & (x >= 17) & (x <= 63)
line_counts = 150.0 * line
four_counts = np.concatenate((true_counts, line_counts[..., None]), axis=-1)
four_spectra = np.vstack((spectra, mneptune))
four_component_scene = simulate_linear_mixture(
    four_counts, four_spectra, detector.channel_centers,
)
measured = add_confocal_noise(
    four_component_scene.image, rng=np.random.default_rng(2026),
)
omitted_fit = SpectralImage(data=measured, axes="YXC").unfold()
pixels = omitted_fit.data
signal_pixels = pixels[pixels.sum(axis=1) > 0]

assumed_components = 3
hull, _ = essential_spectral_pixles(signal_pixels, npc=assumed_components)
starts, _ = unique_profiles(hull.T, nr=assumed_components, f=1.0)
_, fit_spectra, _, fit_info = mcr_als(
    signal_pixels, starts, nIter=200, normL="L1",
    tol_percent=0.01, return_diagnostics=True,
)
fit_counts = positive_least_squares(pixels, fit_spectra, solver="auto")
predicted = omitted_fit.fold_spectra(fit_counts @ fit_spectra).data
omitted_errors = reconstruction_error_maps(measured, predicted)
omitted_pca = residual_pca_from_residuals(omitted_errors.residuals, n_components=3)
print(f"Relative lack of fit: {fit_info.best_relative_l2:.1%}")

The simulated mNeptune line and its location in the RMS, relative L2, and PC1 score maps. A signed residual map and the corresponding PC1 loading curve show which wavelengths contribute.

The overall lack of fit is 18.6%, close to the 17.9% in the original three-component example. The maps show a clearer difference. The line has higher RMS error than the surrounding mixture, and it stands out in the first residual PC score map. The loading curve shows where that pattern varies across wavelengths. The signed residual panel shows measured minus predicted signal at the channel with the largest absolute PC1 loading. At that channel, the model overpredicts signal along the line as it tries to fit the mixture with three components.

On experimental data, a coherent feature like this gives you a region to inspect and a reason to try another fit. The PC loading is a pattern in the unexplained signal. Its shape alone cannot identify the fluorophore or determine how many components to add.

The repository's Python/evaluations/evaluate_dim_line_recovery.py explores the same diagnostic idea with bright spokes, a dim mNeptune line, a broad background, and several blur levels. It also compares four- and five-component fits to test whether adding a component helps recover the line.